Voltammetric determination of hydrogen peroxide content in peracetic acid disinfectant
Loading...
Date
ORCID
item.page.thesis.degree.name
item.page.thesis.degree.level
item.page.thesis.degree.discipline
item.page.thesis.degree.department
item.page.thesis.degree.grantor
item.page.thesis.degree.advisor
item.page.thesis.degree.committeeMember
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Aim of Work. This study presents the development and validation of a voltammetric method for the quantitative determination of hydrogen peroxide (HP) in the peracetic acid-based disinfectant Delakson.
Methodology. The method employs cathodic voltammetry using a carbositall rotating electrode (CRA) as both the working and auxiliary electrode. Measurements were conducted in the potential range of +1.0 to –1.0 V versus a saturated Ag/AgCl/KCl
reference electrode.
Scientific Novelty. The study offers a novel application of cathodic voltammetry with a carbositall rotating electrode for precise HP determination in a commercial disinfectant matrix. The method demonstrated high linearity, sensitivity, and precision, with
low LOD (2.15×10⁻⁵ mol L⁻¹) and LOQ (7.18×10⁻⁵ mol L⁻¹) values, making it a robust analytical tool for real-world applications.
Conclusions. The proposed voltammetric method is reliable, sensitive, and precise for determining hydrogen peroxide in peracetic acid-based disinfectants. It demonstrated excellent reproducibility (RSD values of 0.028, 0.018, and 0.011) and accuracy
(relative errors from –0.77% to +0.92%), with successful application to the Delakson formulation (RSD = 0.012, relative error = +1.69%).
Description
Citation
Voltammetric determination of hydrogen peroxide content in peracetic acid disinfectant / Mozgova О., Blazheyevskiy M., Karpova S., Bondarenko N., Tomarovska T. //
Modern medicine, pharmacy and psychological health. –-Kyiv : Interregional Academy of Personnel Management, 2025. – Issue 2(20). – Р. 204–211.